Docket ID: [NHTSA Docket No. 2007-27133]
SUBJECT CATEGORY: Highway Safety Programs; Proposed Amendments to Model Specifications for Screening; Devices To Measure Alcohol in Bodily Fluids
DOCUMENT SUMMARY: This notice proposes revisions to Model Specifications for Screening Devices to Measure Alcohol in Bodily Fluids (Model Specifications) published in the Federal Register on August 2, 1994 (59 FR 39382). These devices test for the presence of alcohol using breath or bodily fluids such as saliva. The Model Specifications support State laws that target youthful offenders (i.e., ``zero tolerance'' laws) and the Department of Transportation's regulations on Alcohol Misuse Prevention, and encourage industry efforts to develop new technologies (e.g., nonbreath devices) that measure alcohol content from bodily fluids.
This notice proposes to remove testing of Interpretive Screening Devices (ISDs) and use of the Breath Alcohol Sample Simulator (BASS) device from the Model Specifications. The ISDs do not provide an unambiguous test result, as test results for ISDs are subjective and require interpretation by a test administrator or technician. Because the agency has determined the BASS device is not necessary for inclusion in the Model Specifications, this notice proposes to remove all references to the BASS device.
Additionally, in order to ensure product integrity, this notice proposes guidelines for retesting devices when manufacturers contemplate changes, revisions, or upgrades to alcohol screening devices on the Conforming Products List (CPL).
The proposed revisions to these Model Specifications would not affect devices currently listed on the CPL.
SUMMARY: Model specifications and conforming products list—; Alcohol in bodily fluids screening devices,
As indicated in the Model Specifications published in 1994, the agency will modify and improve the Model Specifications as new data and test procedures become available and will alter the test procedures, as necessary, to meet unique design features of specific devices. Since publication of the Model Specifications, the agency has encountered difficulties ensuring the accuracy of testing ISDs and also has determined the use of the BASS is not necessary for inclusion in the Model Specifications. These events make it necessary to revise the Model Specifications.
The Model Specifications currently allow for evaluation of screening devices that require subjective interpretation of test results by a test administrator or technician. These ISDs differ from devices that provide objective test results, including the use of digital technology or the appearance of lights or marks based on the presence or absence of alcohol. For instance, use of pass/fail lights or enzymes that react with alcohol to produce an unambiguous mark provide objective test results.
The Model Specifications require that interpretive devices be evaluated subjectively under five lighting conditions (fluorescent, incandescent, mercury, sodium and daylight) by a panel of ten novice evaluators who are not color blind. Since publication of the Model Specifications, NHTSA evaluated eight separate ISDs. Of these eight ISD evaluations, none resulted in a successful outcome in the panel test described above. In one evaluation, the device passed the test under all lighting conditions except sodium. This device is no longer manufactured. Although many novice evaluators were able to judge the correct test outcome in the eight ISD evaluations, some could not, even though the manufacturers' instructions were conveyed to the evaluators and all evaluators passed tests to determine their color perception ability. This subjective interpretation of test results does not ensure accuracy and precision required to protect public safety. Due to repeated problems in evaluating ISDs, NHTSA is proposing to remove altogether testing of ISDs from the Model Specifications. Specifically, the agency proposes to update sections 3.2, 4.1 and 4.2, delete sections 4.3 and 4.4, and renumber sections accordingly. In addition, the agency proposes to delete from Appendix A all references to interpretive or color indicator tests.
The Model Specifications currently provide for the use of the Breath Alcohol Sample Simulator (BASS) device for providing alcoholin air test samples. The use of the BASS device is not necessary for inclusion in the Model Specifications because the BASS device is intended for use in testing the sampling efficiency of evidential breath testers. There is no sampling efficiency test in the Model Specifications for alcohol screening devices. The alcoholinair test sample for breath alcohol screening devices is supplied by a calibrating unit. Therefore, the agency proposes to remove section 3.5 and all references to the BASS device from these Model Specifications, and renumber sections accordingly. The agency would also revise section 3.4 to include the updated citation for NHTSA's Model Specifications for Calibrating Units.
The Model Specifications provide procedures to conduct special
investigations and retest a device if information gathered indicates
that a device listed on the CPL is not performing in accordance with
the Model Specifications. The agency proposes the addition of Appendix
B to provide guidance regarding notification and retesting when
manufacturers contemplate revisions to devices listed on the CPL. The proposed Appendix follows the language used in the Model
[[Page 71189]]
Specifications for evidential breath testing devices (58 FR 48705).
Upon notification by a manufacturer of a contemplated change to a device listed on the CPL, NHTSA proposes that it would determine whether retesting is required. Such determination would look at several factors, including the nature and reason for the change, the scope of the change, the effects of the change on the performance of the device, and how the change will be documented for the benefit of the user.
NHTSA would list device revisions and whether retesting was required in the next update to the CPL. Appendix B also would state that NHTSA may retest any device listed on the CPL at any time to determine continued compliance and performance with the Model Specifications. A device found not to perform in accordance with the Model Specifications would be subject to the special investigation procedures discussed below.
This notice proposes no changes to the procedures for the Model Specifications other than those discussed above. This section describes the current procedures. The DOT Volpe National Transportation Systems Center (VNTSC), RTV4F, Kendall Square, Cambridge, MA 02142 tests products manufacturers submit to determine whether the products meet the model specifications. Tests are conducted semiannually, or as necessary. Manufacturers are required to apply to NHTSA for a test date by writing to the Office of Behavioral Safety Research, NTI130, NHTSA, 1200 New Jersey Avenue, SE., Washington, DC 20590. At least 30 days are typically required from the date of notification until the test can be scheduled.
One week prior to the scheduled initiation of the test program, manufacturers must deliver their devices to VNTSC. If the devices are disposable, the manufacturer must deliver at least 300 such devices; if the devices are reusable, the manufacturer must submit only a single device. If a manufacturer of a reusable device wishes to submit a duplicate, backup instrument, it may so do. The manufacturer is responsible for ensuring that the devices operate properly and are packaged correctly. The manufacturer must also deliver the operator's manual (or instructions) and the maintenance manual (if any) that would be supplied or is supplied with the purchase of the device, as well as specifications and drawings fully describing the device and its use. Proprietary information will be respected. (See 49 CFR Part 512, regarding the procedure by which NHTSA will consider claims of confidentiality.)
In addition, the manufacturer must submit a selfcertification, certifying that the manufacturer meets the requirements according to the U.S. Food and Drug Administration (FDA) Good Manufacturing Practices regulations for devices used for medical purposes (21 CFR Part 820), and that the device's label meets the requirements in FDA's Labeling regulations for devices used for medical purposes (21 CFR 809.10), even if the devices are not to be used for medical purposes. See Appendix A to this notice.
The manufacturer has the right to check its device(s) between the time of its arrival at VNTSC and the start of the tests, but will have no access to the device(s) during the tests. Any malfunction of a device resulting in failure to complete any of the tests satisfactorily will result in a determination that the device does not conform to the Model Specifications. If a device is found not to conform to the Model Specifications, it may be resubmitted for the next testing cycle after appropriate corrections have been made. The agency reserves the discretion to determine the appropriateness of any retest.
The agency intends to update and republish the CPL in the Federal Register annually. Republications of the CPL add conforming alcohol screening devices tested since the last CPL republication.
NHTSA will continue to provide notification in the Federal Register when the agency amends the Model Specifications as new data and test procedures become available and will retest devices when necessary.
The NHTSA Office of Behavioral Safety Research is the point of contact for information about acceptance testing and field performance of devices. NHTSA requests that users of alcohol screening devices provide both acceptance and field performance data to the Office of Behavioral Safety Research when such data are available. Information from users will help NHTSA monitor whether alcohol screening devices are performing according to the NHTSA Model Specifications.
If information gathered indicates that a device on the CPL is not performing in accordance with the Model Specifications, NHTSA will direct VNTSC to conduct a special investigation. An investigation may include visits to users and additional tests of the device as obtained from the open market. If the investigation indicates that a device actually sold on the market does not meet the Model Specifications, the manufacturer will be notified that the device may be removed from the CPL. In this event, the manufacturer will have 30 days from the date of notification to reply. Based on the VNTSC investigation and any data provided by the manufacturer, NHTSA will decide whether the device should remain on the CPL. If the device is removed from the CPL, the manufacturer will be permitted to resubmit an improved device to VNTSC for testing when it believes the problems causing its failure have been resolved. Upon resubmission, the manufacturer must submit a statement describing what has been done to overcome the problems that led to failure of the device.
If information gathered indicates that the manufacturer of a device on the CPL does not comply with the requirements in FDA's Good Manufacturing Practices regulations for devices used for medical purposes or that the device's label does not comply with the requirements in FDA's labeling regulations for devices used for medical purposes, NHTSA will investigate the matter in consultation with FDA and will notify the manufacturer that the device may be removed from the CPL. The manufacturer will have 30 days from the date of notification to reply. Based on any data provided by the manufacturer and investigative findings, NHTSA will decide whether the device should remain on the CPL. If the device is removed from the CPL, the manufacturer will be permitted to resubmit a selfcertification, certifying that the manufacturer or its device complies with these FDA requirements when it believes the problems causing its noncompliance have been resolved. Upon resubmission, the manufacturer must submit a statement describing what has been done to overcome the problems that led to noncompliance.
These proposed amendments have been analyzed in accordance with the principles and criteria contained in Executive Order 12612, and it has been determined that there are no federalism implications that warrant the preparation of a federalism assessment.
In accordance with the foregoing, the proposed amendments of the
Model Specifications for Screening Devices to Measure Alcohol in Bodily Fluids, are set forth below.
[[Page 71190]]
Model Specifications for Alcohol Screening Devices
These specifications establish performance criteria and methods for
testing of alcohol screening devices. Alcohol screening devices use
bodily fluids to detect the presence of 0.020 or more BAC (see below)
with sufficient accuracy for screening purposes. These specifications
are intended primarily for use in the conformance testing of alcohol screening devices.
2. Classification
Alcohol screening devices designed for a single use. 2.2 Reusable Alcohol Screening Devices
Alcohol screening devices designed to be reused. 3. Definitions
The intoxicating agent in beverage alcohol, ethyl alcohol or other low molecular weight alcohols including methyl or isopropyl alcohol. 3.2 Alcohol Screening Device
A device that is used to detect the presence of 0.020 or more BAC. The device may measure any bodily fluid for this purpose, but shall provide output in BAC units. Test results must be indicated unambiguously by numerical readout or by other means, such as by the use of lights or by the appearance of a distinctive mark but not by color change.
Grams of alcohol per 100 milliliters of blood or grams of alcohol
per 210 liters of breath in accordance with the Uniform Vehicle Code,
Section 11903(a)(5)\1\ (BrAC is often used to indicate that the
measurement is a breath measurement); or grams of alcohol per 100 milliliters of saliva.
\1\ Available from the National Committee on Uniform Traffic
Laws and Ordinances, 107 S. West Street,
A device that produces an alcoholinair test sample of known concentration and that meets the NHTSA Model Specifications for Calibrating Units (72 FR 34742).
Any bodily fluid capable of being used to estimate alcohol concentration, provided the relationship between such bodily fluid and BAC has been established according to scientifically acceptable standards. Such fluids include but are not limited to blood, exhaled deep lung breath and saliva.
Fluids that have been scientifically accepted as equivalent to bodily fluids for testing purposes, such as aqueous alcohol test solutions on a onetoone basis for blood or saliva.
Testing will be performed according to the instructions that normally accompany the submitted device and under the conditions specified in the tests below.
Perform 40 trials under normal laboratory conditions including 20 trials at 0.008 BAC and 20 trials at 0.032 BAC. Use a calibrating unit for this test for breath devices and preparations of bodily fluids or scientifically acceptable substitutes for nonbreath devices. Perform tests using a VNTSC investigator.
To conform at 0.008 BAC, not more than one positive result. To conform at 0.032 BAC, not more than one nonpositive result. 4.2 Test 2. Blank Reading
Perform 20 trials under normal laboratory conditions at 0.000 BAC. Use nonalcoholic human breath for breath devices and nonalcoholic bodily fluids or scientifically acceptable substitutes for nonbreath devices. Perform tests using a VNTSC investigator.
To conform: No positive results. If the device is capable of
providing a reading of greater than 0.000 BAC and less than 0.020 BAC, not more than one such result.
4.3 Test 3. Cigarette Smoke Interference (Only Breath and Saliva Test Devices)
Perform five trials at 0.000 BAC. Select an alcoholfree person who smokes cigarettes for this test. Ask the person selected to smoke approximately one half of a cigarette. Within one minute after smoking, or after a waiting period specified in the manufacturer's instructions, administer the alcohol screening device test according to the manufacturer's instructions. Then ask the person to smoke another inhalation and repeat the test to produce a total of five trials.
To conform: No positive results.
4.4 Temperature
Test at low and high ambient temperature.
Perform 40 trials at 10 degrees Centigrade (C), including 20 trials at 0.008 BAC and 20 trials at 0.032 BAC. Use a calibrating unit for this test for breath devices and preparations of bodily fluids or scientifically acceptable substitutes for nonbreath devices.
To conform at 0.008 BAC, not more than one positive result. To conform at 0.032 BAC, not more than one nonpositive result. 4.4.2 Test 4.2. High Ambient Temperature
Perform trials of 40 devices at 40 degrees C, including 20 trials at 0.008 BAC and 20 trials at 0.032 BAC. Use a calibrating unit for this test for breath devices and preparations of bodily fluids or scientifically acceptable substitutes for nonbreath devices.
To conform at 0.008 BAC, not more than one positive result. To conform at 0.032 BAC, not more than one nonpositive result. 4.5 Test 5. Vibration
Perform 40 trials, including 20 trials at 0.008 BAC and 20 trials at 0.032 BAC. Use a calibrating unit for this test for breath devices and preparations of bodily fluids or scientifically acceptable substitutes for nonbreath devices.
Mount the screening device on a shake table and vibrate the table
in simple harmonic motion through each of its three major axes, as
specified below. Sweep through each frequency range in 2.5 minutes,
then reverse the sweep to the starting frequency in 2.5 minutes.
Disposable testers may be placed in a suitable box mounted on the shake table. Test after vibration.
Amplitude
Frequency (hertz) (inches, peak to
peak)
10 to 30............................................. 0.30
30 to 60............................................. 0.15
To conform at 0.008 BAC, not more than one positive result. To conform at 0.032 BAC not more than one nonpositive result.
Appendix ALabeling Instructions for Alcohol Screening Devices Intended Use
Provide the intended use including the specimen matrix (e.g. saliva, breath), the assay type (quantitative, semiquantitative), the purpose of performing the assay, and the individual designated to perform the assay.
E.g.: This product is intended for the (quantitative, semi quantitative) determination of alcohol indefine matrix (for e.g., saliva, breath, sweat) to perform screening alcohol assays. [[Page 71191]]
This product is recommended for use by individuals who have been trained in the administration of screening devices.
Provide the principles of the procedure for performing the alcohol screening assay.
E.g.: This product uses (alcohol dehydrogenase, infrared technology, etc.) to perform the test.
Describe the chemical reaction sequence, if applicable.
Reagents: List the concentration, strength, and composition of the reactive ingredients.
List the nonreactive ingredients.
Provide instructions for preparing the reagents, if applicable.
Provide instructions for storing the reagents, if applicable.
Provide any signs of deterioration of the reagents, if applicable.
Provide the reagents' shelf life and opened expiration dating, if applicable.
E.g.: Unopened tests are stable until the date printed on the product container when stored at 2228[deg] C. Opened test must be used at once.
Provide a caution not to use the reagents beyond the expiration dating.
1. List any reagents that may be hazardous such as caustic compounds, sodium azide or other hazardous reagents and instructions for disposal, if applicable.
2. Provide warning to user to treat all samples as potentially infective. Include instructions for handling and disposal of the sample.
Provide instructions for collecting and handling the sample.
Provide criteria for specimen rejection, if applicable. Calibration
Disposable tests are precalibrated. No additional calibration is required.
Reusable (Instrumented) tests require calibration.
Provide information regarding how calibrations are to be conducted, if applicable, including the number and concentration of calibrators, and the frequency of calibration.
Provide instructions for calibration and recalibration.
Provide the criteria for acceptability of calibration. Test Procedure (Disposable)
Provide adequate stepbystep instructions for performing the test and determining the results.
Provide adequate stepbystep instruction for performing the test.
Provide the installation procedures and, if applicable, any special requirements.
Provide the space and ventilation requirements.
Provide the description of the required frequency of equipment maintenance and function checks.
Provide the instructions for any remedial action to be taken when the equipment performs outside of operating range.
Provide any operational precautions and limitations.
Provide instructions for the protection of equipment and
instrumentation from fluctuations or interruptions in electrical
current that could adversely affect test results and reports, if applicable.
Quality Control (QC)
If applicable, the function and stability of the test can be determined by the examination of the procedural ``built in'' controls contained in the product. If these controls are not working, the test is invalid and must be repeated.
If external quality control materials are used, provide number, type, matrix and concentration of the QC materials.
Provide directions for performing quality control procedures.
Provide an adequate description of the remedial action to be taken when the QC results fail to meet the criteria for
Provide directions for interpretation of the results of quality control samples.
Describe how the user obtains the test results, from an instrument readout, printout, etc.
Describe the results in terms of blood alcohol concentration.
Describe what concentration indicates a positive result and what concentration indicates a negative result.
List the substances or factors that may interfere with the test
and cause false results including technical or procedural errors. Dynamic Range
Provide the operating range of the product.
Only devices that meet the precision and accuracy of these Model Specifications will be included on NHTSA's Conforming Products List for alcohol screening devices.
List the substances that have been evaluated with your product
that do or do not interfere at the concentration indicated. References
Provide pertinent bibliography.
List an 800 number the user may contact for further information or technical assistance.
Appendix BGuidelines for Retesting of Modified Screening Devices
Manufacturers contemplating revisions to an alcohol screening
device listed on the Conforming Products List (CPL) are advised that
the revision may affect the status of the device on the CPL. The
manufacturer should inform NHTSA of the contemplated change so that
a judgment can be made whether or not retesting the revised alcohol
screening device is necessary. The following lists the type of
information NHTSA uses in determining the necessity to retest an alcohol screening device, and is provided as guidance to
manufacturers:
If necessary for clarity, drawings of the listed and changed device may also be helpful in the NHTSA's deliberations.
If, upon review of information provided by a manufacturer, it is determined that retesting is not warranted, a statement to that effect will be included in the next scheduled CPL update.
Additionally, NHTSA reserves the right to retest any device on
the open market to determine continued compliance and performance in
accordance with these Model Specifications. Devices found not to
comply with or perform in accordance with the Model Specifications
are subject to the investigation provisions stated above in Section II, Procedures.
(Authority: 23 U.S.C. 403; 49 CFR 1.50; 49 CFR Part 501).
Issued on: December 14, 2007.
Marilena Amoni,
Associate Administrator for the Office of Research and Program Development.
[FR Doc. E724282 Filed 121307; 8:45 am]
BILLING CODE 491059P
FOR FURTHER INFORMATION CONTACT For technical issues: Ms. De Carlo Ciccel, Behavioral Research Division, NTI131, National Highway Traffic Safety Administration, 1200 New Jersey Avenue, SE., Washington, DC 20590; Telephone: (202) 3661694. For legal issues: Ms. Allison Rusnak, Office of Chief Counsel, NCC113, National Highway Traffic Safety Administration, 1200 New Jersey Avenue, SE., Washington, DC 20590; Telephone: (202) 3661834.
14 CFR Part 39 40 CFR Part 52 14 CFR Part 71 33 CFR Part 165 50 CFR Part 679 47 CFR Part 73 26 CFR Part 1 40 CFR Part 180 33 CFR Part 117 50 CFR Part 17 44 CFR Part 67 50 CFR Part 648 14 CFR Part 97 40 CFR Part 63 33 CFR Part 100 50 CFR Part 622 50 CFR Part 660 26 CFR Part 301 44 CFR Part 65 39 CFR Part 111 40 CFR Part 300 6 CFR Part 5 40 CFR Part 271 47 CFR Part 64 40 CFR Parts 52 and 81 50 CFR Part 665 10 CFR Part 50 44 CFR Part 64 49 CFR Part 571 39 CFR Part 3020