SUBJECT CATEGORY: Notice Regarding Revisions to the Laboratory Protocol To Measure the Quantity of Nicotine Contained in Smokeless Tobacco Products Manufactured, Imported, or Packaged in the United States
DOCUMENT SUMMARY: The uniform protocol for the analysis of nicotine, total moisture, and pH in smokeless tobacco products, originally published in the Federal Register in 1999 (64 FR 14086), ``Notice Regarding Requirement for Annual Submission of the Quantity of Nicotine Contained in Smokeless Tobacco Products Manufactured, Imported, or Packaged in the United States,'' and revised in the Federal Register on March 14, 2008 (73 FR 13903), implements the requirement of the Comprehensive Smokeless Tobacco Health Education Act (CSTHEA) of 1986 (15 U.S.C. 4401 et seq., Pub. L. 99252) that each entity manufacturing, packaging, or importing smokeless tobacco products shall annually provide the Secretary of Health and Human Services (HHS) with a specification of the quantity of nicotine contained in each smokeless tobacco product. CDC is republishing the notice published in the Federal Register on March 14, 2008 (73 FR 13903) concerning the revision of the protocol for analysis of nicotine in smokeless tobacco products (hereinafter referred to as ``Protocol'') to (1) make a technical change to correct the date when the first report of information under the revised Protocol is due; (2) solicit public comments concerning a change in the Protocol that increased the volume of water in the pH determination from 10 mL to 20 mL, and (3) solicit public comments concerning the addition of the following commercial smokeless tobacco product categories: Dry snuff portion packs, snus, snus portion packs, and pellet or compressed.
The Protocol as published in the Federal Register on March 14, 2008 (73 FR 13903), remains in effect with the technical correction to the date described below.
Technical change: The language in the March 14, 2008 notice stated that ``The first report of information is due June 30, 2008, with subsequent submissions due by March 31 of each year.'' The first report date of information should be 2009 so that the sentence correctly reads: ``The first report of information is due June 30, 2009, with subsequent submissions due by March 31 of each year.''
SUMMARY: Revisions to the Laboratory Protocol to Measure the Quantity of Nicotine Contained in Smokeless Tobacco Products Manufactured, Imported, or Packaged in the,
After evaluating information that has recently come to the
attention of the Centers for Disease Control and Prevention's Office on
Smoking and Health (OSH) regarding low moisture smokeless tobacco
products packaged in portion pouches, OSH conducted an independent
comparison of pH measurements in a variety of low and high moisture
smokeless tobacco products. The results of this comparison, presented
in Table 1, indicate that there is an acceptable (less than 2%) level
of change in pH values when measurements are taken with 20 mL
deionized, distilled water (Condition B) compared to 10 mL of
deionized, distilled water (Condition A). Increasing the volume of
water in the mixture ensures that the matrix is sufficiently fluid to facilitate ease of measure.
[[Page 35396]]
Table 1.Summary of pH Levels Found in Seven Types of Smokeless Tobacco Products: Plug; Loose Leaf or Scrap; Twist; Dry SnuffLow Moisture/No Pouch;
Dry SnuffLow Moisture/Pouch; Snus; and Moist Snuff
pH \a\
Category Smokeless Tobacco Product Condition A 10 mL \b\ Condition B 20 mL \b\
pH Change % Change
Mean \c\ SD \d\ Mean \c\ SD \d\
Plug........................................ Days O Work Chew.............. 5.06
\b\ n = 1.
\c\ Average pH from four measured intervals.
\d\ Standard Deviation.
\e\ Accurately weighed: 2.000
OSH has determined that these revisions will improve the applicability of the protocol and provide guidance to reporting entities and other interested parties for testing of all currently marketed categories of smokeless tobacco. The change in the volume of liquid in the pH determination facilitates the ease of measure of smokeless tobacco pH for all currently marketed smokeless tobacco categories (i.e., plug, twist, moist snuff, dry snuff, snus, loose leaf, chew, moist snuff in portion pouches, smokeless tobacco compressed into a pellet, and dry snuff in portion pouches). Collection of Information
This proposed amendment does not call for any new collection of
information under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501 3520).
Dated: June 13, 2008.
James D. Seligman,
Chief Information Officer, Centers for Disease Control and Prevention.
Revised Protocol for Analysis of Nicotine, Total Moisture, and pH in Smokeless Tobacco Products
I. Requirements \1\ \2\
A. Reagents \3\
1. Sodium hydroxide (NaOH), 2N.
2. Methyl tbutyl ether (MTBE).
3. () Nicotine (Fluka 72290) >99% purity.4 5
4. Quinoline (Aldrich).
5. Standard pH buffers; 4.01, 7.00, and 10.00.
6. Deionized distilled water.
B. Glassware and Supplies
1. Volumetric flasks, class A.
2. Culture tubes, 25 mm x 200 mm, with Teflonlined screw caps.
3. Pasteur pipettes.
4. Repipettors (10 mL and 50 mL).
5. Linear shaker (configured to hold tubes in horizontal position).6 7
6. Weighing dishes, aluminum.
7. Tefloncoated magnetic stirring bars.
8. Polypropylene containers, 50 mL.
1. Robot Coupe Model RSI 2V Scientific Batch Processor.
2. Capillary gas chromatograph, Hewlett Packard, Model 6890, with split/splitless injector capability, flame ionization detector, and a capillary column (Hewlett Packard HP5, Crosslinked 5% PH ME Siloxane, 30 m length x 0.32 mm ID, film thickness 0.25 or 0.52 [mu]m).
3. Orion Model EA 940 pH meter equipped with Orion 8103 Ross combination pH electrode.
Forcedair oven, Fisher Isotemp[supreg], regulated to 99
1. Detector temperature: 250 [deg]C.
2. Injector temperature: 250 [deg]C.
3. Flow rate at 100 [deg]C1.7 mL/min; with split ratio of 40:1.\10\
4. Injection volume: 2 [mu]l.
5. Column conditions: 110185 [deg]C at 10 [deg]C min1; 185240 [deg]C at 6 [deg]C min1, hold at final temperature for 10 min.
There are ten different categories of commercial smokeless tobacco products:
1. Dry snuff;
[[Page 35397]]
2. Moist (wet) snuff;
3. Moist (wet) snuff portion packs;
4. Plug;
5. Twist;
6. Loose leaf;
7. Dry snuff portion packs;
8. Snus;
9. Snus portion packs; and
10. Pellet or Compressed.
Because of their physical characteristics, some of the ten product categories must be ground (whole or in part) before nicotine, total moisture, and pH analyses can be conducted. The objective of grinding the samples is to obtain a homogeneous sample with particles measuring approximately 4 mm. Grinding to achieve this particle size should take no more than 3 minutes. To ensure proper grinding and an adequate amount of the ground sample for analysis, the minimum sample size of all commercial products to be ground should not be less than 100 grams.
To ensure precision of analyses for nicotine, total moisture, and pH, the samples that require grinding should be ground using a Robot Coupe Model RSI 2V Scientific Batch Processor or its equivalent. This is a variable speed (0 to 3000 RPM) processor. The variable speed motor is required to ensure proper grinding of the tobacco tissues (and in the case of pH determination, the portion pack). Elevated temperatures can result in moisture loss and an underestimated value for moisture content. Hence, care must be taken during grinding to avoid elevated temperatures. The bowl should be cleaned after each grinding to obtain accurate results. Freeze or cryogrinding is also an acceptable grinding method.
1. Dry snuff: Dry snuff samples do not need to be ground since the product is a powder. The sample must be thoroughly mixed before weighing for nicotine, total moisture, and pH analysis.
2. Moist (wet) snuff: Moist (wet) snuff samples do not need to be ground. The sample must be thoroughly mixed before weighing for nicotine, total moisture, and pH analysis.
3. Moist (wet) snuff portion packs: The tobacco contents of the moist (wet) snuff portion packs do not need to be ground for nicotine, total moisture, or pH analysis. The tobacco packaging material (the ``pouch'') should be separated from the tobacco and ground to obtain particles measuring approximately 4 mm for pH analysis. The tobacco of the moist (wet) snuff portion pack and the ground pouch are combined and thoroughly mixed before pH analysis.
4. Plug tobacco: Break or cut apart plugs and add in portions to grinder at 2000 RPM. Reduce RPM or stop grinding if sample bowl becomes warm. Pulse the Robot Coupe, when needed, to complete grinding. Grind samples until approximately 4 mm in size. The total grinding time should be no more than 3 minutes.
5. Twist tobacco: Separate twists, add to grinder and grind at 2000 RPM. Reduce RPM or stop grinding if sample bowl becomes warm. Continue grinding until sample particles are approximately 4 mm in size. The total time for grinding should be no more than 3 minutes.
6. Loose leaf: Grind in the same manner as described in 4 and 5 to obtain product with particle size of approximately 4 mm.
7. Dry snuff portion packs: The tobacco contents of the dry snuff portion packs do not need to be ground for nicotine, total moisture, or pH analysis. The tobacco packaging material (the ``pouch'') should be separated from the tobacco and ground to obtain particles measuring approximately 4 mm for pH analysis. The tobacco of the dry snuff portion pack and the ground pouch are combined and thoroughly mixed before pH analysis.
8. Snus: Snus samples do not need to be ground since the product is a powder. The sample must be thoroughly mixed before weighing for nicotine, total moisture, and pH analysis.
9. Snus portion packs: The tobacco contents of the snus portion packs do not need to be ground for nicotine, total moisture, or pH analysis. The tobacco packaging material (the ``pouch'') should be separated from the tobacco and ground to obtain particles measuring approximately 4 mm for pH analysis. The tobacco of the snus portion pack and the ground pouch are combined and thoroughly mixed before pH analysis.
10. Pellet or compressed: Break apart compressed tobacco pellets
and add in portions to grinder at 2000 RPM. Reduce RPM or stop grinding
if sample bowl becomes warm. Pulse the Robot Coupe, when needed, to
complete grinding. Grind samples until approximately 4 mm in size. The total grinding time should be no more than 3 minutes.
II. Nicotine Analysis \12\
A. Calibration Standards
Weigh 10.00 grams of quinoline, transfer to a 250 mL volumetric flask and dilute to volume with MTBE. This solution will be used for calibration of the instrument for the nicotine calibration curve (II.A.2), for the standards addition assay (II.B), and for preparation of the extracting solution (II.D).
a. Weigh 1.0000 gram of nicotine into a clean, dry 100 mL volumetric flask and dilute to volume with MTBE. This gives a nicotine concentration of 10 mg/mL for the stock solution.
b. Accurately pipette 0.5 mL of IS from stock solution (II.A.1) to five clean, dry 50 mL volumetric flasks. To prepare a nicotine standard corresponding to a concentration of 0.8 mg/mL, pipette exactly 4.0 mL of the nicotine standard (II.A.2.a) to a 50 mL volumetric flask containing the internal standard and dilute to volume with MTBE. To obtain nicotine concentrations equivalent to 0.6, 0.4, 0.2, and 0.1 mg/ mL, pipette precisely 3.0, 2.0, 1.0, and 0.5 mL, respectively, of the nicotine standard into the four remaining flasks and dilute to volume with MTBE.
c. Transfer aliquots of the five standards to auto sampler vials and determine the detector response for each standard using gas chromatographic conditions described in I.E.
d. Calculate least squares line for linear equation from these
standards by obtaining the ratio of Area
[GRAPHIC] [TIFF OMITTED] TN23JN08.029
Where:
X = Concentration of nicotine in mg
Y = Area
a = intercept on the ordinate (y axis)
The final result will be reported in the following units:
Concentration of nicotine = mg of nicotine/gram of tobacco sample.
e. Determine the recovery of nicotine by pipetting 10 mL of the 0.4
mg/mL nicotine standard to a screw capped tube containing 1.0 mL of 2 N
NaOH. Cap the tube. Shake the contents vigorously and allow the phases
to separate. Transfer an aliquot of the organic phase to an injection
vial and inject. Calculate the concentration of nicotine using the
equation of the line in II.A.2.d above. This should be repeated two
more times to obtain an average of the three values. The recovery of nicotine can be obtained by using the following equation:
[GRAPHIC] [TIFF OMITTED] TN23JN08.030
[[Page 35398]]
Prior to analyzing a smokeless tobacco product for nicotine content, the testing facility must validate the system to verify that matrix bias is not occurring during nicotine extraction. This is done by analyzing the nicotine calibration standards in the same vegetable matrix as the smokeless tobacco. The first time each smokeless tobacco product is tested and whenever a change is made to the product formulation (including a change to the tobacco blend or cultivar), the Standards Addition Assay will be performed, and documentation of its performance and of the nicotine concentrations selected for the standard curve (II.B.2) will be submitted to the Centers for Disease Control and Prevention.
1. Using an analytical balance, accurately weigh 1.000
2. Prepare a fivepoint standard curve for the Standards Addition
Assay. The standard curve must consist of nicotine concentrations that
encompass the range of values expected from adding known concentrations
of the nicotine standard (II.A.2.a) to a measured quantity of the
smokeless tobacco product (1.000
3. Pipette 5 mL of 2 N NaOH into each tube. Cap each tube. Swirl to wet sample and allow to stand 15 minutes.\13\
4. Pipette 50 mL of extraction solution (II.D.1) into each tube. Cap each tube and tighten.\14\
5. Place tubes in rack(s), place racks in linear shaker in horizontal position and shake for two hours.
6. Remove rack(s) from shaker and place in vertical position to allow the phases to separate.
7. Allow the solvent and nicotine supplemented samples and the blank to separate (maximum 2 hours).
8. Transfer aliquots of the five standards and the blank from the extraction tubes to sample vials and determine the detector response for each using gas chromatographic conditions described in I.E.
9. Subtract the Area
10. Calculate least squares line for linear equation from the corrected standards as described above (Equation 1) in II.A.2.d. The final corrected result will be reported in the following units: Concentration of nicotine = mg of nicotine/gram of tobacco sample.
11. Determine the recovery of nicotine by pipetting 10 mL of the
0.4 mg/mL nicotine standard to a screw capped tube containing 1.0 mL of
2 N NaOH and 10 mL of extraction solution (II.D.1). Cap the tube and
tighten. Shake the contents vigorously and allow the phases to
separate. Transfer an aliquot of the organic phase to an injection vial
and inject. Calculate the concentration of nicotine using the equation
of the line above in II.A.2.d. This should be repeated two more times
to obtain an average of the three values. The recovery of nicotine can be obtained by using Equation 2: Recovery =
Nicotine
12. Compare the results of steps II.A.2 and II.B. If they differ by a factor of 10% or more, the recovery of nicotine from the aqueous matrix is not equivalent to recovery from the vegetable matrix of the smokeless tobacco product. In this instance, the nicotine concentration of the smokeless tobacco product must be determined from a nicotine calibration curve prepared from nicotine standards in a vegetablebased matrix.
At least two quality control pools at the high and low ends of the expected nicotine values are recommended to be included in each analytical run. The pools should be analyzed in duplicate in every run. The quality control pools should be available in sufficient quantity to last for all analyses of a product.
1. Extraction solution is prepared by pipetting 10 mL of the IS from the stock solution (II.A.1) to a 1000 mL volumetric flask and diluting to volume with MTBE.
2. Using an analytical balance, accurately weigh 1.000
3. Pipette 5 mL of 2 N NaOH into the tube. Cap the tube. Swirl to wet sample and allow to stand 15 minutes.\13\
4. Pipette 50 mL of extraction solution into tube, cap tube and tighten.\14\
5. Place tubes in rack(s), place racks in linear shaker in horizontal position and shake for two hours.
6. Remove rack(s) from shaker and place in vertical position to allow the phases to separate.
7. Allow the solvent and sample to separate (maximum 2 hours). Transfer an aliquot from the extraction tube to a sample vial and cap.
8. Analyze the extract using GC conditions as described above (I.E) and calculate the concentration of nicotine using the linear calibration equation. Correct percent nicotine values for both recovery and weight of sample by using Equation 3.\17\
9. Report the final nicotine determination as mg of nicotine per gram of the tobacco product (mg nicotine/gram), to an accuracy level of two decimal places for each brand name (e.g., Skoal Bandits Wintergreen, Skoal Long Cut Cherry, Skoal Long Cut Wintergreen, etc.). All data should include the mean value with a 95% confidence interval, the range of values, the number of samples tested, the number of lots per brand name, and the estimated precision of the mean. Information will be reported for each manufacturer and variety (including brand families and brand variations) and brand name (e.g., Skoal Bandits Wintergreen, Skoal Long Cut Cherry, Skoal Long Cut Wintergreen, etc.). [[Page 35399]]
A. This procedure is a modification of AOAC Method 966.02 (1990)
and is referred to as ``Total Moisture Determination'' because it
determines water and tobacco constituents that are volatile at temperatures of 99
B. Accurately weigh 5.00 grams of the sample (ground to pass <= 4 mm screen) \20\ into a weighed moisture dish and place uncovered dish in oven.\21\ Sample each smokeless tobacco brand name according to the provided testing frequency schedule.\19\ The number of products sampled should reflect an acceptable level of precision.\16\ The test material is to be representative of the product that is sold to the public and therefore should consist of sealed, packaged samples of finished product that is ready for commercial distribution. Samples are to be analyzed in duplicate.
C. Do not exceed 1 sample/10 sq in (650 sq cm) shelf space, and use
only 1 shelf. Dry 3 hr at 99
D. Report the final moisture determination as a percentage (%), to an accuracy level of one decimal place for each brand name (e.g., Skoal Bandits Wintergreen, Skoal Long Cut Cherry, Skoal Long Cut Wintergreen, etc.). All data should include the mean value with a 95% confidence interval, the range of values, the number of samples tested, the number of lots per brand name, and the estimated precision of the mean. Information will be reported for each manufacturer and variety (including brand families and brand variations) and brand name (e.g., Skoal Bandits Wintergreen, Skoal Long Cut Cherry, Skoal Long Cut Wintergreen, etc.).
A. Test samples as soon as possible after they are received. Sample each smokeless tobacco brand name according to the provided testing frequency schedule.19 The number of products sampled should reflect an acceptable level of precision.16 The test material is to be representative of the product that is sold to the public and therefore should consist of sealed, packaged samples of finished product that is ready for commercial distribution. Samples are to be analyzed in duplicate.
B. Accurately weigh 2.00 grams of the sample. Place in a 50 mL polypropylene container with 20 mL deionized distilled water.
C. Place Tefloncoated magnetic stirring bar in container and stir mixture continuously throughout testing.
D. Measure pH of sample after a twopoint calibration of the pH meter to an accuracy of two decimal places using standard pH buffers (4.01 and 7.00 or 7.00 and 10.00) that will encompass the expected pH value of the smokeless tobacco product.
E. The first time pH values are determined for a smokeless tobacco product, measure the pH of the smokeless tobacco product at 5, 15, and 30 minutes. If there is no systematic variation in pH values with time, all subsequent pH determinations are made at 5 minutes. If there is systematic variation in pH values, continue to measure the pH of the smokeless tobacco product until the pH value is stable and does not vary more than 10% over 15 minutes. Report the final pH value.
F. Report the final pH determination to an accuracy level of two decimal places for each brand name (e.g., Skoal Bandits Wintergreen, Skoal Long Cut Cherry, Skoal Long Cut Wintergreen, etc.). All data should include the mean value with a 95% confidence interval, the range of values, the number of samples tested, the number of lots per brand name, and the estimated precision of the mean. Information will be reported for each manufacturer and variety (including brand families and brand variations) and brand name (e.g., Skoal Bandits Wintergreen, Skoal Long Cut Cherry, Skoal Long Cut Wintergreen, etc.).
G. Estimate the unionized (free) nicotine content with the
HendersonHassel Balch equation (Equation 4), based on measured pH and nicotine content.
[GRAPHIC] [TIFF OMITTED] TN23JN08.032
[GRAPHIC] [TIFF OMITTED] TN23JN08.033
pKa = 8.02 (CRC Handbook of Chemistry and Physics, 19891990) [B] = amount of unionized (free) nicotine
[BH+] = amount of ionized nicotine
H. Report the final estimated unionized (free) nicotine as a
percentage (%) of the total nicotine content, to an accuracy level of
two decimal places and as mg of unionized (free) nicotine per gram of
the tobacco product (mg unionized (free) nicotine/gram), to an
accuracy level of two decimal places for each brand name (e.g., Skoal
Bandits Wintergreen, Skoal Long Cut Cherry, Skoal Long Cut Wintergreen,
etc.). All data should include the mean value with a 95% confidence
interval, the range of values, the number of samples tested, the number
of lots per brand name, and the estimated precision of the mean.
Information will be reported for each manufacturer and variety
(including brand families and brand variations) and brand name (e.g.,
Skoal Bandits Wintergreen, Skoal Long Cut Cherry, Skoal Long Cut Wintergreen, etc.).
Sample calculation:
Mean total nicotine = 10.30 (mg/g)
Mean pH = 7.50
pKa = 8.02
[GRAPHIC] [TIFF OMITTED] TN23JN08.034
[GRAPHIC] [TIFF OMITTED] TN23JN08.035
[[Page 35400]]
[GRAPHIC] [TIFF OMITTED] TN23JN08.036
[GRAPHIC] [TIFF OMITTED] TN23JN08.037
[GRAPHIC] [TIFF OMITTED] TN23JN08.038
[GRAPHIC] [TIFF OMITTED] TN23JN08.039
[GRAPHIC] [TIFF OMITTED] TN23JN08.040
[GRAPHIC] [TIFF OMITTED] TN23JN08.041
V. Assay Criteria for Quality Assurance
All quality control parameters must be determined within the laboratory in which they are to be used. At least 10 withinlaboratory runs must be performed to establish temporary confidence intervals for the quality control parameters. Permanent limits should be established after 20 runs and should be reestablished after each additional 20 runs.
The coefficient of determination between Areanicotine/AreaIS and nicotine concentration should be equal to 0.99 or higher. Any calibration standard having an estimated concentration computed from the regression equation (Equation 1) which is different from its actual concentration by a factor of 10% can be excluded from the calibration curve. Up to two concentrations may be excluded, but caution should be used in eliminating values, since bias may be increased in the calibration curve. If an outlier value is eliminated, its duplicate value must also be discarded to avoid producing a new bias. All unknowns must fall within the calibration curve; therefore, duplicate values excluded at either end of the calibration curve will restrict the useful range of the assay.
The mean estimated nicotine concentration in a pool should be compared with the established limits for that pool based on at least 20 consecutive runs. An analytical run should be accepted or rejected based upon the following set of rules adapted from Westgard et al. (1981).
1. When the mean of one QC pool exceeds the limit of x
2. When the mean nicotine concentrations in two QC pools in the same run exceed the same direction, then the run must be rejected. The same direction is the condition in which both pools exceed either the x + 2 SD or the x 2 SD limits.
3. When the mean nicotine concentrations in one or two QC pools
exceed their x
4. When the mean nicotine concentrations in two QC pools are different by more than a total of 4 SD, then the run must be rejected. This condition may occur, for example, when one QC pool is 2 SD greater than the mean, and another is 2 SD less than the mean.
The comments and notes listed below can be described as Good
Laboratory Practice guidelines; they are described in detail in this
protocol to ensure minimal interlaboratory variability in the
determination of nicotine, total moisture, and pH in smokeless tobacco.
\1\ This protocol assumes that the testing facility will
implement and maintain a stringent Quality Assurance/Quality Control
program to include, but not be limited to, regular interlaboratory
comparisons, determination of the quality and purity of purchased
products, and proper storage and handling of all reagents and samples.
[[Page 35401]]
\2\ When a specific product or instrument is listed, it is the
product or instrument that was used in the development of this
method. Equivalent products or instruments may also be used. Use of
trade names is for identification only and does not constitute
endorsement by the Public Health Service or the U.S. Department of Health and Human Services.
\3\ All chemicals, solvents, and gases are to be of the highest purity.
\4\ Companies must ensure that the purity of the nicotine base
is certified by the vendor and that the chemical is properly stored.
However, nicotine base oxidizes with storage, as reflected by the
liquid turning brown. If oxidation has occurred, the nicotine base
should be distilled prior to use in making a standard solution.
\5\ A suggested method for the determination of nicotine purity is CORESTA Recommended Method No. 39.
\6\ Horizontal shaking will allow more intimate contact of this
three phase extraction. There is a minimal dead volume in the tube due to the large sample size and extraction volume. This
necessitates horizontal shaking.
\7\ If a linear shaker is not available, a wrist action shaker
using 250 mL stoppered Erlenmeyer flasks can be substituted. Values
for nicotine are equivalent to those obtained from the linear shaker.
\8\ After installing a new column, condition the column by
injecting a tobacco sample extract on the column, using the
described column conditions. Injections should be repeated until
areas of IS and nicotine are reproducible. This will require
approximately four injections. Recondition column when instrument
has been used infrequently and after replacing glass liner.
\9\ Glass liner and septum should be replaced after every 100 injections.
\10\ Most older instruments operate at constant pressure. To
reduce confusion, it is suggested that the carrier gas flow through the column be measured at the initial column temperature.
\11\ The testing facility must ensure that samples are obtained
through the use of a survey design protocol for sampling ``at one
point in time'' at the factory or warehouse. The survey design
protocol must address short, medium, and longterm smokeless
tobacco product variability (e.g., variability over time and from
container to container of the tobacco product) in a manner
equivalent to that described for cigarette sampling in Annex C of
ISO Protocol 8243. Information accompanying results for each sample should include, but not be limited to:
For each productmanufacturer and variety (including brand families and brand variations) and brand name (e.g., Skoal Bandits, Skoal Long Cut Cherry, Skoal Long Cut Wintergreen, etc.):
1. Product ``category,'' e.g., loose leaf, plug, twist, dry snuff, moist (wet) snuff, etc.
2. Lot number.
3. Lot size.
4. Number of randomly sampled, sealed, packaged (so as to be representative of the product that is sold to the public) smokeless tobacco products selected (sampling fraction) for nicotine, moisture, and pH determination.
5. Documentation of method used for random sample selection.
6. ``Age'' of product when received by testing facility and storage conditions prior to analysis.
\12\ Extraction of nicotine and pH determination must be
performed with reagents and samples at a room temperature of 2225
[deg]C. Room temperature should not vary more than 1[deg]C during extraction of nicotine or pH determination.
\13\ Use nonglass 10 mL repipette for transferring NaOH solution.
\14\ Use 50 mL repipette for transferring MTBE.
\15\ For dry snuff, use 0.500
\17\ When analyzing new smokeless tobacco products, extract
product without IS to determine if any components coelute with the
IS or impurities in the IS. This interference could artificially lower calculated values for nicotine.
\18\ The calculated nicotine values for all samples must fall
within the low and high nicotine values used for the calibration
curve. If not, prepare a fresh nicotine standard solution and an
appropriate series of standard nicotine dilutions. Determine the
detector response for each standard using chromatographic conditions described in I.E.
\19\ The testing frequency for each smokeless tobacco brand name
(e.g., Skoal Bandits Wintergreen, Skoal Long Cut Cherry, Skoal Long
Cut Wintergreen, etc.) is based on the manufacturing duration (refer
to table below). Each smokeless tobacco brand name will be sampled
and tested for nicotine, total moisture, and pH no fewer than twice and no more than four times during a calendar year.
Test
Manufacturing duration in weeks frequency
*
up to and including 4........................................ 2
up to and including 28....................................... 3
up to and including 52....................................... 4
* Use a statistical program to determine random sampling dates based on
the total manufacturing duration during a calendar year. Sampling
dates should fall on actual manufacturing days for the product when
test material that is representative of the product that is sold to
the public (consisting of sealed, packaged samples) is available. If a
statistically determined sampling date falls on a day that does not
meet this criterion, sample the product on the next date that does meet the criteria.
For smokeless tobacco brand names with episodic production during a calendar year, the total number of sampling dates is determined by the sum of the individual test frequencies, not to exceed four. For the purpose of the Protocol, episodic production is defined as manufacturing intervals separated by periods of 30 or more days when the smokeless tobacco brand name is not manufactured.
Example 1: Within a single calendar year a smokeless tobacco brand name is manufactured from January 1 to March 31 and from September 1 to December 15. The testing frequency for the first manufacturing interval is 3 and for the second manufacturing interval is 3. The Protocol allows that each smokeless tobacco brand name be tested for nicotine, total moisture, and pH no more than four times during a calendar year. Therefore, 4 random sampling dates, as described in the footnote to the above table, are determined for the smokeless tobacco brand name. The values for nicotine, moisture, and pH determinations, and unionized (free) nicotine calculations and the mean of the 4 data points for that smokeless tobacco brand name are reported.
Example 2: Within a single calendar year a smokeless tobacco brand name is manufactured from April 5 to May 3 and from September 1 to December 15. The testing frequency for the first manufacturing interval is 2 and for the second manufacturing interval is 3. The values for nicotine, moisture, and pH determinations, and unionized (free) nicotine calculations and the mean of the 4 data points for that smokeless tobacco brand name are reported.
Example 3: Within a single calendar year a smokeless tobacco
brand name is manufactured from January 1 to January 15 and from
September 1 to September 22. The testing frequency for the first
manufacturing interval is 2 and for the second manufacturing
interval is 2. Four random sampling dates are selected to fall
within the 6 weeks of manufacturing for the smokeless tobacco brand
name. The values for nicotine, moisture, and pH determinations, and
unionized (free) nicotine calculations and the mean of the 4 data points for that smokeless tobacco brand name are reported.
\20\ The method is a modification of AOAC Method 966.02 (1990)
in that the ground tobacco passes through a 4 mm screen rather than a 1 mm screen.
\21\ When drying samples, do not dry different products (e.g.,
moist (wet) snuff, dry snuff, loose leaf) in the oven at the same
time since this will produce errors in the moisture determinations.
\22\ The method is a modification of a method published by Henningfield et al. (1995).
AOAC (Association of Official Analytical Chemists). Official Methods of Analysis. 966.02: Moisture in Tobacco. (1990) Fifth Edition. K. Helrich (ed). Association of Official Analytical Chemists, Inc., Suite 400, 2200 Wilson Boulevard, Arlington, Virginia 22201 USA.
CORESTA (Centre de Coop[eacute]ration pour les Recherches Scientifiques relatives au Tabac). Recommended Method No. 39: Determination of the purity of nicotine and nicotine salts by gravimetric analysisTungstosilic acid method. November 1994. 87 90.
CRC Handbook of Chemistry and Physics. R.C. Weast, D.R. Lide, M.J. Astle, and W.H. Beyer (eds). 70th ed. Boca Raton, Florida: CRC Press (19891990) D162.
Henningfield, J.E., Radzius A., Cone E.J. (1995). Estimation of available nicotine content of six smokeless tobacco products. Tobacco Control 4:5761.
ISO (International Organization for Standardization). IOS 8243: CigarettesSampling. (1991). Second Edition. Prepared by Technical Committee ISO/TC 126, Tobacco and tobacco products. International Organization for Standardization, Case Postale 56, CH1211 Genve 20, Switzerland.
Westgard, J.O., Barry P., Hunt M., and Groth T. (1981). A multi
rule Shewhart chart for quality control in clinical chemistry. Clinical Chemistry 27:493.
[FR Doc. E814112 Filed 62008; 8:45 am]
BILLING CODE 416318P
FOR FURTHER INFORMATION CONTACT Matthew McKenna, M.D. Director, Office on Smoking and Health, Telephone: (770) 4885701.
14 CFR Part 39 40 CFR Part 52 14 CFR Part 71 33 CFR Part 165 50 CFR Part 679 26 CFR Part 1 40 CFR Part 180 47 CFR Part 73 50 CFR Part 17 33 CFR Part 117 44 CFR Part 67 50 CFR Part 648 14 CFR Part 97 33 CFR Part 100 40 CFR Part 63 26 CFR Part 301 50 CFR Part 622 39 CFR Part 111 40 CFR Part 300 44 CFR Part 65 50 CFR Part 660 40 CFR Part 271 40 CFR Parts 52 and 81 47 CFR Part 64 50 CFR Part 665 49 CFR Part 571 44 CFR Part 64 21 CFR Part 522 14 CFR Part 23 47 CFR Part 76